QP1680 TOC/TNb Analyzer – High‑Temperature Lab Analyzer with 65‑Position Autoloader

QP1680 TOC/TNb Analyzer – High‑Temperature Lab Analyzer with 65‑Position Autoloader
Brand: HACH
Product Code: QP1680
Reviews: (0)
Pre-Order
0 UZS
Buy in one click
Specifications: (see all)
Measured Parameters
TOC; TC; TIC; NPOC; TNb
Measurement Method
High‑temperature catalytic combustion (up to 1050 °C), CO₂ detection by NDIR; TNb by chemiluminescence
Measurement Range
0–30,000 mg/L C for TOC/TIC/TC/NPOC and TNb
Detection Limit
TOC/TIC/TC/NPOC: 50 µg/L; TNb: 20 µg/L
Measurement Accuracy
Repeatability: up to 10 mg/L < 5%; > 10 mg/L < 2%
Delivery Information DEPENDING ON THE SIZE AND QUANTITY OF THE PRODUCT Payment Pay quickly and conveniently

The Hach QP1680 Laboratory Analyzer enables simultaneous measurement of Total Organic Carbon (TOC), Total Inorganic Carbon (TIC implicit), Total Carbon (TC), and Total Nitrogen (TNb) using high‑temperature catalytic combustion (up to 1050 °C). It includes a 65‑position liquid auto sampler with stirrer, supports particle injection up to 800 µm, and features intuitive software for data analysis and system diagnostics. The analyzer delivers a wide range analysis up to 30,000 mg/L C, with TOC detection limits around 50 µg/L, analysis cycle times of ~3–4 minutes, and repeatability <5 % below 10 mg/L and <2 % above. Compact footprint, high reliability — ideal for QA/QC and environmental laboratories.

Tags: TOC Analyzer

TOC Analyzer
Measured Parameters
TOC; TC; TIC; NPOC; TNb
Measurement Method
High‑temperature catalytic combustion (up to 1050 °C), CO₂ detection by NDIR; TNb by chemiluminescence
Measurement Range
0–30,000 mg/L C for TOC/TIC/TC/NPOC and TNb
Detection Limit
TOC/TIC/TC/NPOC: 50 µg/L; TNb: 20 µg/L
Measurement Accuracy
Repeatability: up to 10 mg/L < 5%; > 10 mg/L < 2%
Analysis Time
Approximately 3–4 minutes per sample
Throughput
Up to ~15–20 samples per hour (depending on cycle time)
Sample Type
Liquid samples including particulate up to 800 µm
Sample Volume
10–1000 µL
Power Supply
100–240 VAC ±10%, 50/60 Hz; analyzer power max ~750 W
Interfaces and Outputs
Modbus RTU/TCP, optional Profibus, 4–20 mA, digital outputs
Software / PC Compatibility
Intuitive software, data analysis, diagnostics; compatible with Hach Claros portal
Display Type
No onboard display; controlled via PC/software/Hach Claros
Dimensions
700 × 440 × 380 mm (H×W×D)
Weight
≈ 32 kg
Country of Origin
USA