Water analysis
Measured / controlled parameters
pH, ORP (mV), conductivity, total dissolved solids (TDS), salinity, resistivity, dissolved oxygen (DO), oxygen saturation, BOD, ion-selective measurements (ammonium, ammonia, chloride, fluoride, nitrate, sodium), temperature
Measurement / analysis range
pH: 0–14 pH; mV: −1500…1500 mV; Conductivity: 0.01 µS/cm…200.0 mS/cm; TDS: 0.00 mg/L…50.0 g/L; Salinity: 0–42 ppt; Resistivity: 2.5 Ω·cm…49 MΩ·cm; DO: 0.1–20.0 mg/L, 1–200% saturation; Temperature: −10…+110 °C
Measurement accuracy / precision
pH: ±0.002 pH; mV: ±0.1 mV; Conductivity: ±0.5% FS; DO: ±0.1 mg/L; Salinity: ±0.1%
Temperature measurement range
−10…+110 °C
Temperature compensation
Automatic compensation for pH, conductivity, DO (barometric)
Analysis method / operating principle
Electrochemical methods using replaceable electrodes and parameter-specific sensors
Measurable concentration range of substances
Depends on the ion-selective electrode used (ammonium, chloride, fluoride, nitrate, sodium, etc.)
Analysis / measurement time
Instant real-time measurement display
Cuvette / electrode / sensor type
pH: PHC281; Conductivity/TDS/Salinity: CDC401; DO: LDO or LBOD probes; ISE: respective sensor
Number of measurement channels
2 channels (Multi/2 Channels)
Availability of automatic mode / programmable cycles
Auto-stabilization (“press to read”), automatic data saving, interval measurement selection
Sample volume / sample volume range
—
Housing / working part material
Housing with IP54 protection (resistant to water splashes, dust-proof)
Housing protection class (IP)
IP54
Interfaces and outputs
USB Type A and Type B for PC, printer, keyboard, flash drive
Software / PC compatibility
HQD Series Meter Data Transfer Utility
Memory / data storage availability
500 records / FIFO, automatic storage with calibration data
Display / indication type
Graphic LCD display 440×160 px with backlight
Calibration
pH: 1–3 points; DO: 0–100% or mg/L with 0-point; ISE: 2–5 points; ORP: by ORP standards
Light source / detector type
—
Power supply
4 × AA batteries, AC 100–240 V power adapter
Operating conditions (temperature, humidity)
0…+60 °C, humidity up to 90% (non-condensing)
Dimensions
859×175×235 mm
Weight
750 g (without batteries), 850 g (with batteries)
Scope of supply
HQ440D meter, PHC281 electrode, CDC401 cell, pH buffers 4.01/7.00/10.01, electrode storage solution, NaCl standard, stand with electrode holder, power adapter, USB cables, user manual
Certificates and compliance with standards
CE, WEEE
Additional functions
Printer connectivity, automatic data transfer to PC or USB, custom buffer sets, calibration alerts and reminders